Panel Meter
MODEL:451B


MODEL:451B

■ Thermocouple input/Platinum-resistance thermometer input
■ High-resolution of 0.1 or 0.01°C display
■ Water- and dust-proof IP65 front panel

Model Designation

451B
 
1
2
3

1.Power supply voltage
Code Power source voltage
A
AC100V~240V
B
DC12~24V
C
DC110V

2. Data output 1
Code Data output Output impedance Tolerable Load resistive
Blank
-
09
Analog output DC0~10V
(Available scaling,Default:1~5V )
0.1Ω or less 500Ω or more(DC1~5V)
1kΩ or more(DC0~10V)
29
Analog output DC0~20mA
(Available scaling,Default:4~20mA)
5MΩ or more 0~600Ω(DC4~20mA)
BP
BCD output(TTL level positive logic)
-
BN
BCD output(TTL level negative logic)
-
DP
BCD output(transistor output,source type)
-
DN
BCD output(transistor output,sink type)
-
E0
RS-232C
-
E1
RS-485
-

3. Data output 2
(Available- 09 and -29 of Data output 1 only)
Code Data output
Blank No outout
E0
RS-232C
E1
RS-485

Measuring Input
Thermocouple
Sensor
Measuring range Display range Accuracy
R
100.0~1700.0℃ -50.0~1800℃ ±(0.1% of rdg + 0.6digit) [100.0~500.0℃]
±(0.1% of rdg + 0.5digit) [500.0~1700.0℃]
K
-100.0~1300.0℃ -200.0~1400.0℃ ±(0.1% of rdg + 0.6digit) [-100.0~0.0℃]
±(0.1% of rdg + 0.5digit) [0.0~1300.0℃]
E
-130.0~1000.0℃ -250.0~1050.0℃ ±(0.1% of rdg + 0.5digit)
J
-140.0~1200.0℃ -210.0~1250.0℃
T
-200.0~400.0℃ -250.0~420.0℃
B
600.0~1800.0℃ -20.0~1820.0℃ ±(0.1% of rdg + 0.6digit)
N
-100.0~1300.0℃ -230.0~1350.0℃ ±(0.1% of rdg + 0.5digit)

Resistance thermobulb

Sensor
Measuring range Display range Accuracy
Pt100Ω
range[1]
-200.0~850.0℃ -200.0~870.0℃ ±(0.1% of rdg + 0.2digit) [0.0~100.0℃]
±(0.2% of rdg + 0.3digit) [-200.0~0.0℃] [100.0~850.0℃]
 
Pt100Ω
range[2]
-150.00~150.00℃ -180.00~180.00℃ ±(0.1% of rdg + 0.2digit) [0.0~100.0℃]
±(0.2% of rdg + 0.3digit) [-150.0~0.0℃] [100.0~150.0℃]
JPt100Ω -200.0~645.0℃ -200.0~660.0℃ ±(0.1% of rdg + 0.2digit) [0.0~100.0℃]
±(0.2% of rdg + 0.3digit) [-200.0~0.0℃] [100.0~645.0℃]


Basic Specifications

Display 0~99999, red or green LED (character height 15.2mm), “-” polarity, with zero-suppress function.
During the measuring mode, the display switches from current value to peak memory, bottom memory,
display amplitude, and current value, by pushing PB key.
Current value, PB 1 second, Memory, Peak memory, PM illuminates, Bottom memory, BM illuminates, Display amplitude
Unit ( ℃/ oF) Selectable by the parameter setting ( oF = ℃ x 9/5+32)
Over-range indication Blinking with minimum or maximum value of display range
Resolution RTD input: 0.1 ℃ (0.01℃ when PT100Ω range 2)
Thermocouple input 0.1 ℃
External resistance 500Ω Max. for Thermocouple input
Wire resistance 5Ω Max. per wire for Thermocouple input
Overload DC +/- 10V
Sampling rate Approx. 5 times / sec.
Display cycle selectable from 200ms, 400ms, 1s, 2s, 4s, 5s
Response within 2 samplings or within display cycle, whichever longer
Input method Single ended, floating input
A/D conversion ⊿-∑ conversion method
Noise Rejection Normalmode (NMR) - 50dB or more.
Common mode (CMR) - 110dB or more.
Noise through power supply line - 1000V (at AC voltage power supply)
Insulation Resistance -DC500V 100MΩor more.
Withstanding Voltage Input terminals - Case : AC1500V each for 1 min.
Power supply terminals - Case : AC1500V each for 1 min.
Power supply terminals - Input and output terminals : AC500V each for 1min.
Input terminals - Output terminals : AC500V each for 1 min.
Power Supply AC 100 to 240V (90-250V) 50/60HZ, DC12 to 24V (9-32V)
Power Consumption 451B itself: Approx. 7VA at100V AC, 9VA at 200VAC, 300mA at 12V. 150mA at 24VDC.
Operating Temperature 0 to 50℃
Storage Temperatur e -20 to 70℃
Weight Approx. 300g
Housing protection IP65 for the front panel, IP20 for the rear case, IP00 for terminals
Mounting Method Panel mount with the bracket


Operation Manual