Digital Meter Relay
MODEL 452G


Voltage and Current input meter Relay for High-speed, Highprecision Measurement

MODEL 452G

■ DC2CH input, four arithmetic operations function
■ High-speed sampling 2000 times/sec.
■ Respond to the rapid input changes by comparison function against the previous data
■ Ideal for mass-production workstation by the synchronous measurement function

Model Designation

452G 1 1 1 1 1 1
 
1
2
3
4
5
6

1. Measuring input (Two input are the same and same range)


Code Measuring range Input impedance Accuracy of display Over load
03
±999.9mV 100MΩ ±(0.1% of F.S + 1digit)
for two input
±250V
04
±9.999V 1MΩ
V2
0~5V 1MΩ
09
1~5V 1MΩ
19
4~20mA 12.4Ω ±150mA


2. Power supply voltage

Code Power source voltage
A
AC100V~240V
B
DC12~24V


3. Power source for sensor(Floating output)

Code output voltage output current
Blank
-
2
DC5V±10% 100mA
3
DC12V±5% 150mA *1
5
DC24V±5% 100mA *2
NOTE: *1 100mA for the DC power supply
*2 50mA for the DC power supply


4. Data output 1

Code Data output Output impedance Tolerable Load resistive
Blank
-
09
Analog output DC0~10V
(Available scaling,Default:1~5V )
0.1Ω or less 1kΩ or more
29B
Analog output DC0~20mA
(Available scaling,Default:4~20mA)
5MΩ or more 0~500Ω
BP
BCD output(TTL level positive logic)
-
BN
BCD output(TTL level negative logic)
-
DP
BCD output(transistor output,source type)
-
DN
BCD output(transistor output,sink type)
-
E0
RS-232C
-
E1
RS-485
-
EC
Decimal point external control
-


5. Data output 2 (Available- 09 and -29 of Data output 1 only)

Code Data output
Blank No outout
E0
RS-232C
E1
RS-485
EC
Decimal point external control


6. Comparoson output

Code specifications
TN
Open collector output (NPN)
RY
Relay output



Basic Specifications

Input 2CH input (operation result is Y between A and B), A and B is not isolated
Dual input operation Y=A+B, Y=A-B, Y=A×B, Y=A/B, Y=(A×B)/10, Y=(A×B)/100, Y=(A×B)/1000, But Ymax=99999
Main Display 0~9999, “-” polarity, with zero-suppress function. +/- 99999 when calibrating
PV: red or green LED (character height 15.2mm)
Indication color: RR, RG, GR, GG (R: Red when any AL1 to AL4 turn ON, G: Green when allALturn off.)
Sub-display SV1 and SV2: red LED (character height 7.6mm)
Display Display data output, current value, peak memory, bottom memory, and display amplitude
“PM” lights up at the peak memory display
“BM” lights up at the bottom memory display.
Both “PM” and “BM” lights up at the amplitude display
Scaling Full-scale: -99999 to +9999
Offset: -99999 to +9999
Display cycle Selectable from 67ms, 400ms, 1s, 2s, 4s, 5s
Resolution 1/10000
Sampling rate 2000 times / sec.
Response 2ms at 1CH, 4ms at 2CH
Input method Single ended, floating input
A/D conversion Successive approximation method
Noise Rejection Normalmode (NMR) - 40dB or more.
Common mode (CMR) - 110dB or more.
Noise through power supply line - 1000V (at AC voltage power supply)
Insulation Resistance DC500V 100MΩ or more.
Withstanding Voltage Input terminals - Case : AC2000V each for 1min.
Power supply terminals - Case : AC2000V each for 1 min.
Power supply terminals - Input and output terminals : AC1500V each for 1min
Input terminals - Output terminals : AC500V each for 1 min.
Power Supply AC 100 to 240V 50/60HZ, DC12 to 24V, DC110V
AllowablePowerRange AC 90 to 250V, DC9 to 32V, DC100 to 170V
Power Consumption 452G itself Approx. 8.5VA at100V AC, 10.5VA at 200VAC, 300mA at 12V. 150mA at 24VDC.
Power supply unit: Approx. 7VA at100V AC, 9VA at 200VAC, 200mA at 12V. 100mA at 24VDC.
Operating Temperature 0 to 50 ℃
Storage Temperature -20 to 70 ℃
Weight Approx. 300g (60g of the power supply unit)
Housing protection IP65 for the front panel, IP20 for the rear case, IP00 for terminals

 

Operation Manual